0086-531-88190005
Company NewsIndustry News
Dyonics Equipment for ME300BF HCXO300 Power Source with High Precision Short Arc Continuous Xenon La
Source:SHANDONG LASER SOURCE    Release time:2022-07-26
Dyonics Equipment for ME300BF HCXO300 Power Source with High Precision Short Arc Continuous Xenon Lamp Power Supply 201 - 300w

Product Features

1. Simple Remote/Automatic transfer function, Use own jump line can realize Remote/ Automatic model  transition. 2. Output current can be adjusted from DC11A to DC24A. 3. Perfect Protection Function,short-circuit protection、Overheating protection and overload protection,increased the power supply reliability. 4. Constant current and trigger module integration design,output connect the xenon lamp directly,easy to use. 5. Output is power limitation of constant current mode. 6. Active PFC,reducing  input harmonic interference,Increase the power factor,Input voltage is suitable for the global network voltage.

Application:

medical laparoscopy, gastroscope, endoscope, xenon lamp cold light source

Product Pictures



Specification
 




Input
Voltage 90-265VAC
Frequency 50Hz/60Hz
Power Factor >0.90
Cold Surge Current ≤42A(AC220Vinput) ≤20A(AC110Vinput)
Current Leakage ≤0.3mA(AC240Vinput) ≤0.15mA(AC120Vinput)




Output
Max Current DC24A
Trigger voltage 14VDC±5%
Current Ripple Ipp≤5%
Stability (Start half an hour) ≤±0.5%
auxiliary output 14VDC±2% 1A
Remote/Automatic Control Remote: Using external switch to control power up/down;
Automatic:Power Supply on after connect the AC power .
Efficiency ≥80%(full load)
Mean Time Between Failure (MTBF) ≥50000H
Power on delay time ≤ 1s
Withstand Voltage Input-Shell: AC1500V ≤5mA 60S;
Protection Output over-current protection, short circuit protection, power limit, over-temperature protection
Environment Operating Temperature:(0~50)℃, Relative Humidity (RH)≤90%
Storage tmperature -10-85℃
Cooling Way Force-Air Cooling (200-400LFM)
PCB Dimension(mm) L×W×H=171×114×80

External Interface

1、Installation Size Picture:

2、External Interface Definition:
1)L、N、E connect AC input L、N、Ground respectively.
2)P1 is Aux output terminals, output is 12V1A,positive and negative terminalsas P1.
3)P2 is Remote control terminal, external switch can control HCXO300-S output current up/down.
4)P3is Remote/Automatic switch terminal, Change the jumper position can be realized,as the P1.
5)LAM POS,LAM NEG is output terminal, LAM POS connect xenon lamp positive, LAM NEG connect the xenon lamp negative.


Precautions

1.Note the Anode and cathode output--before connect to Xenon Lamp, please confirm the xenon lamp and xenon lamp power source anode and cathode are Corresponding, reverse prohibited, in order to avoid damage the xenon lamp. 2.When Xenon lamp turns off, please wait three minutes to restart, to protect lamp electrodes and extend working life of xenon lamp.
3.Ventilation around the xenon lamp power supply should ensure the smooth flow of ventilation fan at the non-blocking.
4.If the xenon lamp power supply placed in another bigger case, we must ensure that case air convection, so as to play a cooling effect, otherwise easily lead to overheating of the chassis.
5.Remote power control interface and output circuit isolation measures have been adopted, need not regard to isolate problems when wiring it.

 

Warning

The output terminal with high voltage, please don’t measure the output voltage directly!!!
Shandong Laser Source Technology Co.,Ltd
No.15 Lanxiang Road, Tianqiao Area, Jinan city 250032, Shandong China
Sale Hotline:                 0086-531-88190005
Technical Consultant: 0086-531-88190006
Production engineers:0086-531-58988699

CONTACT US

Copyright©2015 Shandong Laser Source Technology Co.,Ltd All Rights Reserved